1
GATE EE 1999
MCQ (Single Correct Answer)
+2
-0.6
For a dual $$ADC$$ type $$3\,\,{1 \over 2}$$ digit $$DVM$$, the reference voltage is $$100mV$$ and the first integration time is set to $$300$$$$ms$$. For some input voltage, the ''deintegration'' period is $$370.2ms$$. The $$DVM$$ will indicate
A
$$123.4$$
B
$$199.9$$
C
$$100.0$$
D
$$1.414$$
2
GATE EE 1999
MCQ (Single Correct Answer)
+2
-0.6
The voltage phasor of a circuit is $$10\angle {15^ \circ }V$$ and the current phasor is $$2\angle - {45^ \circ }A.$$ The active and the reactive powers in the circuit are
A
$$10$$ $$W$$ and $$17.32$$ $$V$$ Ar.
B
$$5$$ $$W$$ and $$8.66$$ $$V$$ Ar.
C
$$20$$ $$W$$ and $$60$$ $$V$$ Ar.
D
$$20\sqrt 2 $$ and $$10\sqrt 2 V$$ Ar.
3
GATE EE 1999
MCQ (Single Correct Answer)
+1
-0.3
Electro dynamic type watt meters have large errors while measuring power in a.c circuits at low power factor conditions, since the voltage across and the current though the
A
current coil are not in phase
B
current coil are not in quadrature
C
pressures coil are not in phase
D
pressure coil are not in quadrature.
4
GATE EE 1999
MCQ (Single Correct Answer)
+2
-0.6
Two $$100$$ $$\mu A$$ full scale $$PMMC$$ meters are employed to construct a $$10$$ $$V$$ and a $$100$$ $$V$$ full scale voltmeter. These meters will have figures of merit (sensitivities) as
A
$$10\,k\Omega /V$$ and $$10\,k\Omega /V$$
B
$$100\,k\Omega /V$$ and $$10\,k\Omega /V$$
C
$$10\,k\Omega /V$$ and $$100\,k\Omega /V$$
D
$$10\,k\Omega /V$$ and $$1\,k\Omega /V$$